The NeoPrep System normalizes libraries to 10 nM. Use the following table to determine insert, concentration, and load size for your system and the resulting expected cluster density.
System | Insert Size | Predenatured Concentration (nM) | Denatured/Diluted Intermediate Concentration (pM) | Pooled Denatured Samples Load (pM) | Expected Cluster Density (mm2) |
---|---|---|---|---|---|
MiSeq | 350 | 4 | 40 | 10–14 | 600–1000 |
550 | 4 | 40 | 12-16 | 600–1000 | |
NextSeq | 350 | 4 | 40 | 2-3 | ~200 |
550 | 4 | 40 | 2-3 | ~200 | |
HiSeq 2500 TruSeq v4 | 350 | 4 | 40 | 18-22 | 600-1000 |
550 | 4 | 40 | 20-24 | 600-1000 | |
HiSeq 2500 v3 | 350 | 4 | 40 | 18-22 | 600-1000 |
550 | 4 | 40 | 20-24 | 600-1000 | |
HiSeq 2500 Rapid Run (v1) | 350 | 4 | 40 | 12-16 | 600-1000 |
550 | 4 | 40 | 14-20 | 600-1000 |